Products

Classification

Company Culture    Fluid Power    Hydraulic Components    Digital Pressure Gauge    AMETEK Crystal 15KPSIXP2I-FWV (System F) Complete Pressure Calibration System

AMETEK Crystal 15KPSIXP2I-FWV (System F) Complete Pressure Calibration System

AMETEK Crystal System F (FWV) complete pressure calibration system, with an XP2i digital gauge (0–15000 psi / 1000 bar), T-1-CPF hydraulic hand pump, high-pressure hose and custom case, water medium, intrinsically safe.
online consultation

Product Details

1. Product Overview

15KPSIXP2I-FWV is the AMETEK Crystal System F high-pressure complete pressure calibration system (Pump System). It combines an XP2i digital gauge (0–15000 psi / 1000 bar) with a T-1-CPF high-pressure hydraulic hand pump in a custom case, ready to use out of the box for calibrating high-pressure gauges, switches and transmitters in the field and lab.

Per Crystal's system table, System F is the 0–15000 psi / 1000 bar high-pressure system with a T-1-CPF hand pump; within this system FWV is the distilled-water version and includes a high-pressure hose to extend reach during calibration (FOV is the oil version). The kit contains the XP2i gauge, T-1 pump, high-pressure hose, MPF15K/MPF-1/4BSPF fittings, bonded seals and a custom carrying case.

The XP2i gauge has a die-cast nickel-plated marine-grade aluminum case, IP66/IP67, intrinsic safety (ATEX/IECEx/CSA), −10 to +50°C compensation and ~562 g. The whole kit is a portable solution for high-pressure hydraulic, oil/gas wellhead and process high-pressure calibration.

2. Specifications

表格

Item Specification
Manufacturer AMETEK / Crystal Engineering
Order number 15KPSIXP2I-FWV
Product type Complete pressure calibration system (System F)
Pressure range 0–15000 psi / 1000 bar
Gauge Crystal XP2i digital gauge (15KPSI)
Pressure source T-1-CPF high-pressure hydraulic hand pump
Medium Distilled water (FWV; FOV is oil)
Hose High-pressure rated hose (extended reach)
Accessories MPF15K / MPF-1/4BSPF fittings, bonded seals, custom case
Gauge accuracy Standard 0.1% of reading (S2 ±0.02%FS / S5 ±0.05%FS options)
Gauge enclosure Die-cast nickel-plated marine-grade aluminum, IP66/IP67
Hazardous-area cert. Intrinsic safety: ATEX / IECEx / CSA
Gauge temp. range −10 to +50 °C
Gauge weight ~562 g (with batteries)

5. Model Information

  • Order number: 15KPSIXP2I-FWV
  • Product description: Crystal System F complete calibration system, XP2i 15000psi gauge + T-1 hydraulic hand pump, water medium, with high-pressure hose and custom case.
  • Material number note: "15KPSI" = XP2i gauge range 15000 psi (~1000 bar, the XP2i top); "XP2i" = gauge model; "F" = System F high-pressure pump system; "W" = distilled-water medium (FOV is oil); "V" = includes high-pressure hose. This is a complete system (gauge + pump + hose + case), not a standalone gauge. For a bare gauge or the oil FOV version, please email us.
  • Series features:
    • 0–15000 psi / 1000 bar high-pressure system
    • XP2i gauge + T-1-CPF hand pump, out of the box
    • Distilled-water medium with high-pressure hose
    • Intrinsically safe ATEX/IECEx/CSA, IP66/IP67
    • Custom case for field/lab calibration
  • Related products in series: System FOV (oil version); Systems A/B/C/E/G/H; XP2i/XP3i bare gauges; GaugeCalHP comparator
  • Matching products: CPF fittings and hoses, distilled water/hydraulic oil, FastCalXP software, carry case

6. Why Choose Us

  1. Exact model match: We verify supply of 15KPSIXP2I-FWV as the System F water-based complete system (gauge + pump + hose + case), distinguishing it from oil FOV.
  2. Authorized channel supply: AMETEK Crystal calibration systems through regular channels, with gauge calibration data.
  3. Selection support: We help confirm range, water/oil medium, and whether S2/S5 accuracy or DL logging is needed.
  4. Delivery & after-sales: Quotation, lead-time confirmation and order follow-up provided; for questions, please email us.

7. FAQ

Q1: Is 15KPSIXP2I-FWV a single gauge?

A: No — it is a System F complete calibration system: XP2i 15000psi gauge + T-1-CPF hand pump + high-pressure hose + custom case.

Q2: What is the range?

A: 0–15000 psi ≈ 1000 bar, the XP2i range top.

Q3: FWV vs FOV?

A: Both are System F high-pressure systems; FWV is distilled-water medium with a high-pressure hose, FOV is the oil version.

Q4: What medium?

A: Distilled water (FWV). Do not mix media to avoid contamination.

Q5: Accuracy?

A: XP2i standard is 0.1% of reading; for higher accuracy choose S2 (±0.02%FS) or S5 (±0.05%FS) — please email us.

Q6: Can it be used in hazardous areas?

A: The XP2i gauge is intrinsically safe with ATEX/IECEx/CSA and IP66/IP67, suited to high-pressure hydraulic and oil/gas field calibration.